Abstract
In the paper the hemodynamics in the blood vessel and in the bypass graft is simulated and analyzed. A new construction of a bypass graft with a cuff is applied. The goal of the research is evaluation of influence of the ratio between the length and the height of the bypass graft cuff upon the structure of the blood flow and pathogenesis of the blood vessel. The results of the analysis indicate that blood flow velocity at a junction of the bypass graft and the blood vessel depends on the parameters of the cuff. The best results are obtained when the said ratio is 1.25 and 1.5 due to uneven distribution of pressure in the cuff. Choosing such parameters of a bypass cuff ensures better hemodynamics and absence of haemostasia, thus reducing a risk of formation of thrombi in the bypass graft and the blood vessel.
About this article
Received
March 19, 2011
Accepted
May 14, 2012
Published
June 30, 2012
Keywords
blood vessels
bypass end-to-end
hemodynamics
FEM
Copyright © 2012 Vibroengineering
This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.